BME dashboard
187 devices · 12 open work orders · 3 PPM due this week
Critical: AC-003 sterilizer fault · WO-2026-0150 · Steris field engineer dispatched ETA 2h. Open WO
Total devices
187
across 11 departments
Active
178
95% uptime
Open work orders
12
2 critical · 3 high
Devices in fault
2
AC-003 · BME-MON-0014
Work orders by kind
12 open
Preventive PPM
6
Corrective fault
3
Calibration
2
Commissioning
1
PPM compliance (last 30d)
96%
PPMs scheduled28
Completed on time27
Overdue1
Cost (₹)2,84,500
Avg MTTR (faults)8.2h
Recent work orders
View all
CodeDeviceKindSeverityStatusDueAssignee
active 178service 7fault 2
Asset codeModelUDIManufacturerDepartmentStatusNext PPMNext cal.
Drager Evita V300
BME-VEN-0007 · Ventilator
activeICU
UDI01094390070112
SerialEV3-23-0712
ManufacturerDrager
Safety classCF (IEC 60601)
Current locationICU · Bay 4
Commissioned2023-07-12
Warranty until2027-07-11
VendorDrager India
Maintenance schedule
PPM frequencyQuarterly
Last PPM2026-03-15 · pass
Next PPM due2026-06-15 (27d)
Last calibration2026-03-01 · pass
Next calibration2026-09-01 (105d)
Faults this year0
Maintenance timeline
PPM completed — pass
2026-03-15 · Rajan Gupta · vendor: Drager India · cost ₹12,500
Calibration completed — pass
2026-03-01 · cert: drager_cal_2026_03_01.pdf
PPM completed — pass
2025-12-15 · vendor: Drager India
Fault — battery low warning
2025-11-22 · resolved next day · battery replaced
Commissioning complete
2023-07-12 · Drager engineer Mr. Sharma
All 12 Open 6 In progress 3 Awaiting parts 2 Completed (30d) 18
CodeDeviceTitleKindSeverityStatusDueCost
Critical · awaiting parts. Steris field engineer dispatched. Estimated parts arrival 16:00. Last BI test recorded as fail at 14:18.
Work order
CodeWO-2026-0150
DeviceAC-003 · Steris Amsco V-PRO maX
Kindcorrective_fault
Severitycritical
Statusawaiting_parts
Reported byRita Sharma · CSSD
Reported at2026-05-18 14:22
Due (SLA)2026-05-22 18:00 (4d)
AssigneeRajan Gupta + Steris field eng
Linked incidentINC-2026-0045
Cost (estimated)₹86,500
Steps
Step 1 · Visual inspection
14:30 · Rajan Gupta · vacuum pump seal degraded
Step 2 · Diagnostic run
14:50 · pre-vacuum test fails · code E-411
Step 3 · Vendor contact
15:10 · Steris dispatch confirmed · parts ETA 16:00
Step 4 · Parts replacement
pending · vacuum pump kit + door gasket
Step 5 · Re-test (BI + Bowie-Dick)
pending
Step 6 · Sign off & release to operations
pending
Comments & updates
Rita Sharma · 14:22"Cycle aborted mid-vacuum. Sterilizer is making unusual hissing sound."
Rajan Gupta · 14:30"Confirmed vacuum pump seal failure. Logged code E-411."
Rajan Gupta · 15:10"Steris field engineer Mr. Patel en route. Parts confirmed in transit."
Due dateDeviceModelDepartmentFrequencyVendorStatus
2026-05-22AC-003Steris Amsco V-PRO maXCSSDMonthlySterisoverdue · in WO
2026-05-30AC-002Belimed WD340 SteamFlowCSSDMonthlyBelimeddue in 11d
2026-06-12AC-001Steris Amsco V-PRO maXCSSDMonthlySterisscheduled
2026-06-15BME-VEN-0007Drager Evita V300ICUQuarterlyDrager Indiascheduled
2026-06-20WD-001Getinge 86-2 Washer-DisCSSDMonthlyGetingescheduled
2026-07-22BME-DEF-0003Philips HeartStart MRxOTBi-annualPhilipsscheduled
DateDeviceResultPerformed byVendorCertificateNext due
2026-05-12BME-DEF-0003 · Philips HeartStartpassedPhilips engPhilipscert.pdf2026-10-12
2026-05-08AC-001 · Amsco V-PRO maXpassedSteris engSteriscert.pdf2026-08-30
2026-04-30BME-INF-0021 · Infusomat SpaceadjustedB.Braun engB.Brauncert.pdf2027-01-10
2026-03-15BME-VEN-0007 · Evita V300passedDrager engDrager Indiacert.pdf2026-09-01
2026-02-22US-001 · Sonorex DT255HpassedRajan Guptain-housecert.pdf2026-11-22